Calibration, the Blank and What Makes a Measurement Trustworthy · graine 1 · A4, économe en encre. Le corrigé s'imprime sur sa propre page, pour les adultes.
En anglais

From instrument signals to trusted numbers

Sciences · Chimie · de 18 à 19 ans
Nom ______________________   Date ____________
  1. What does the blank correct for?

    • Background signal present without the analyte
    • Shifts in how the analyte itself responds
    • Mistakes in weighing the sample
  2. Why must calibration standards bracket the sample?

    • So the line looks longer on the graph
    • So the sample is read between standards, not beyond them
    • So every standard gets used at least once
  3. You may delete an outlier just because it spoils a neat line.

    En entourer un :   Vrai   Faux

  4. For that seawater metal measurement, what should the blank be?

    • Pure analyte with no water at all
    • The same seawater matrix minus the analyte
    • An empty dry instrument reading
  5. A metal must be measured in seawater. Which calibration approach fits?

    • External calibration with pure water standards
    • Internal standard with a clean solvent
    • Standard addition into the seawater itself
  6. When should you choose an internal standard over external calibration?

    • When sample is lost unevenly during preparation
    • When the samples are perfectly clean waters
    • When only one standard exists
  7. Can the blank correct for a matrix effect? Explain your verdict.

    • No, it cannot fix how the matrix changes the response
    • Yes, blanks fix every kind of error
    • Yes, if you measure the blank twice
  8. A calibration line looks perfect, yet results run consistently high. What is the likely cause?

    • The blank was measured twice
    • The instrument needs recalibrating daily
    • A matrix effect biases every result
LightMySky · lightmysky.comW1-mt_UNkQFonlWJ-s1

Corrigé

Pour les adultes. Replier cette page avant de passer le reste.

From instrument signals to trusted numbers W1-mt_UNkQFonlWJ-s1

  1. Background signal present without the analyte · Blanks subtract background, not broken responses.
  2. So the sample is read between standards, not beyond them · Reading between standards is interpolation you can trust.
  3. Faux · Deletions need proof of a concrete mistake, stated openly.
  4. The same seawater matrix minus the analyte · Blanks mimic everything except the thing measured.
  5. Standard addition into the seawater itself · Adding standards to the sample carries the matrix along.
  6. When sample is lost unevenly during preparation · The internal companion suffers the same losses and corrects them.
  7. No, it cannot fix how the matrix changes the response · Background subtraction and response shifts are different jobs.
  8. A matrix effect biases every result · Good looking lines can still sit on shifted responses.