The Beer-Lambert Law and Quantitative Absorption
Absorbance is proportional to concentration and path length, which makes a spectrophotometer a concentration meter. The proportionality fails in predictable ways, and knowing where it fails is what keeps the numbers honest.
What a learner can do afterwards
- Uses the law to find a concentration from an absorbance and a calibration line
- Chooses the analytical wavelength from a spectrum and explains why the peak is chosen
- Lists the reasons the law fails at high concentration and with stray light
- Converts between transmittance and absorbance and explains the logarithmic relation
1 · Read
Absorbance is proportional to concentration and path length, which makes a spectrophotometer a concentration meter. The law reads A equals epsilon times l times c, where epsilon scores how eagerly the chemical soaks up that colour. Once you know epsilon, flip the equation to read any unknown concentration straight off its absorbance. A solution of 0.01 moles per liter in a 1 cm tube with absorbance 2 gives epsilon 200.
Choose the peak of the spectrum for analysis. There the chemical drinks in that wavelength most eagerly, so even tiny amounts move the reading. A high epsilon means a sensitive test, while a low one needs plenty of chemical before absorbance stirs.
Transmittance and absorbance tell one story on two scales. Transmittance is the fraction of light getting through, and absorbance is minus log of that fraction. Full light through means absorbance 0, a tenth through means 1, and a hundredth means 2. Each absorbance unit passes ten times less light, which is the logarithmic link in action.
Respect where the law fails. At high concentration the straight line bends, and stray light reaching the detector bends it further. A sample above the top standard cannot be trusted, so dilute it into range and remeasure. Never extend the line beyond tested standards.
Absorbance scales with amount, logs link it to light, and the line only rules its tested range.
More to read
How to Calculate Molar Absorptivity
Molar absorptivity, also called the molar extinction coefficient, tells you how strongly a chemical absorbs light at a particular wavelength. Think of it as a score for how eager the molecule is to soak up that color of light. A high score means even a tiny amount of the chemical blocks a lot of light. A low score means you need a lot of it before the absorbance changes much.The Beer-Lambert law connects everything: A = epsilon x l x c. A is the absorbance, l is the path length the light travels, and c is the concentration. Epsilon is the molar absorptivity you want to find.To calculate it, rearrange the equation to epsilon = A divided by (l x c). First measure the absorbance of a solution whose concentration you already know. Then divide that absorbance by the path length times the concentration. For example, if a solution with a concentration of 0.01 moles per liter in a 1 centimeter tube gives an absorbance of 2, then epsilon = 2 divided by (1 x 0.01) = 200 liters per mole per centimeter.This number is useful because once you know epsilon for a chemical, you can flip the equation around and use any absorbance reading to find an unknown concentration.
2 · Watch
Take it off screen
Where it sits
Learn first
Optional connections
This opens up
8 questions wait behind this lesson, each with its answer explained. Every answer feeds the sky: stars light as they are learned, and dim when it is time to come back.